Original Article
Year : 2017   |  Volume : 5   |  Issue : 2   |   Page : 14-19  

Stability indicating reproducible high performance liquid chromatographic method for determination of losartan potassium and atenolol in bulk and tablet dosage forms

P. S. Tungenwar, Sufiyan Ahmad, V. M. Shastry, Tabrej Mujawar

DOI: 10.31555/jpbs/2017/5/2/14-19

Correspondence Address:Department of Quality Assurance, Gangamai College of Pharmacy, Nagaon, Dhule, Maharashtra, India

Source of Support: None, Conflict of Interest: Nil


DOI: 10.4103/2231-4040.197331

Abstract  

An accurate, precise and reproducible high-performance liquid chromatographic (RP-HPLC) method was developed for quantitative estimation of losartan potassium and atenolol simultaneously in tablet dosage forms. Young Lin (S.K.) gradient system ultraviolet detector and RP C18 (Thermo) with 250 mm × 4.6 mm i.d. and 5 μm particle size. Methanol 0.1% O-phosphoric acid (65:35) was used as the mobile phase for the method. The detection wavelength was 274 nm, and flow rate was 0.9 ml/min. In the developed method, the retention time of losartan potassium and atenolol was found to be 7.76 min and 4.05 min, respectively. The drug was subjected to oxidation, acid hydrolysis, alkaline hydrolysis, and heat to apply stress condition for degradation. The method was validated for specificity, linearity, precision, accuracy, robustness, and solution stability. The linearity, precision, range, and robustness were within the limits as specified by the International Conference on Harmonization Guidelines. Hence, the method was found to be simple, accurate, precise, economic, and reproducible. Hence, the proposed methods can be used for the routine quality control analysis of losartan potassium and atenolol in bulk drug as well as in formulations.

Keywords: Atenolol, degradation, losartan potassium, reproducible high performance liquid chromatographic, stress condition

How to cite this article:
Tungenwar PS, Ahmad S, Shastry VM, Mujawar T. Stability indicating reproducible high performance liquid chromatographic method for determination of losartan potassium and atenolol in bulk and tablet dosage forms. J Pharm BioSci 2017;5(2):14-19.

Introduction

Losar tan potassium (LK) is monopotassium salt of 4-butyl-4- chloro-1-[[2′-(1H-tetrazol-5-yl) [1, 1 ′-biphenyl]-4- yl] methyl]- 1H-imidazole-5-methanol (Figure 1)[1,2] and shows angiotensin II receptor antagonist used as an antihypertensive.[3] Atenolol (ATN) is (RS)-4-(2-hydroxy-3-isopropylaminopropoxy) phenylacetamide (Figure 2),[4] which shows beta adrenergic blocker used for hypertension. This reduces the volume of the blood, decreasing blood return to the heart and thus cardiac output and, by other mechanisms, is believed to lower peripheral vascular resistance. Literature survey reveals the availability of few methods for estimation of both losartan K and atenolol includes ultraviolet (UV), HPLC as alone or in combination with other drugs.[5-10] No method has been reported for stability indicating method development for simultaneous estimation of both drugs in binary dosage form. This work emphasizes on the stability indicating assay method development for simultaneous estimation of losartan K and atenolol in their combined dosage form by RP-HPLC.
The purpose of this study was to develop a stability indicating method for the simultaneous determination of atenolol and losartan K in bulk drugs and to apply the developed method for the quantitative determination of these drugs from tablets. The HPLC technique was chosen because of its previously mentioned advantages. The proposed method was able to separate the compounds of interest and their degradation products within 10 min. Thereafter, this method was validated as per International Conference on Harmonization (ICH) Guidelines.[11,12] A literature survey has shown that a stability indicating HPLC method for the simultaneous determination of atenolol and losartan K has not been developed. The previously developed methods have been able to separate both the drugs during a minimum run time, but they were not stability indicating, i.e., the separation of various degradation products, employing ICH prescribed stress conditions, was not achieved. In this research article, we report the development and validation of a stability indicating LC method for the simultaneous determination of atenolol and losartan K in pharmaceutical dosage form. It separates drug components from degradation products under ICH suggested stress conditions (acid and base hydrolysis, oxidation, photolysis, and thermal stress).[13-15]

Materials and Methods

Instruments

The analysis of the drug was carried out on Young Lin (S.K.) gradient system UV detector. Equipped with RP C18 (Thermo), (4.6 mm × 250 mm; 5 μm), SP930 D pump, a 20 μl injection loop and UV730D absorbance detector and running autochro-3000 software.

Materials and reagents

Atenolol and losartan K were obtained as gift samples from Cipla Ltd., Mumbai and Micro Labs Ltd Bangalore, respectively. O-phosphoric acid and methanol were HPLC grade procured from Space Lab. Nashik, hydrochloric acid, sodium hydroxide, and hydrogen peroxide (AR) grade procured from Loba Chemie. Nashik. The pharmaceutical preparations of binary combination of atenolol and losartan K that is Losar Beta tablet (Unichem Laboratories Ltd.). The commercial formulation of atenolol and losartan K is available in the ratio of 1:1 (50/50 mg) in tablet.

Chromatographic conditions

RP C18 (Thermo), (4.6 mm × 250 mm) particle size packing 5 μm; detection wavelength 274 nm; flow rate 0.9 ml/min; temperature ambient; sample size 20 μl; mobile phase methanol: Water (OPA 0.05%) (65:35).

Preparation of standard stock solution

Stock solution (1000 μg/ml)

Stock solutions were prepared by weighing 10 mg each of LK and Ate. The weighed drugs were transferred to two separate 10 ml volumetric flasks and adjust the volume with mobile phase. The stock solutions were sonicated for 10 min. Then, filtered through Whatman’s filter paper.

Standard solutions (100 μg/ml)

Take 1 ml solution from stock solution and transfer separately in 10 ml volumetric flask and dilute volume up to 10 ml with mobile phase.

The HPLC analysis was performed on reversed phase high performance liquid chromatographic system with isocratic elution mode using a mobile phase of methanol: O-phosphoric acid (65:35 v/v) on RP C-18 column (250×4.6 mm, 5 μm particle size) with 0.9 ml/min flow rate at 274 nm using UV detector (Figure 3).

Forced degradation study

According to ICH Guideline, the limit of degradation approximately 0-20%. The amount of degradation given in percentage (Table 1).

Acid hydrolysis

These samples were prepared by weighing the drugs 10 mg each and transfer in 10 ml volumetric flask. Then, addition of 2 ml HPLC grade methanol in it and dissolve completely. After complete dissolution add hydrochloric acid (0.1 N) and adjust volume 10 ml with it. After complete preparation of the solution, store it at 800c about 3 h in a water bath (Figure 4).


0fa7a377-fd99-4988-b36a-a8c60aaa7ca8.png

Alkaline hydrolysis

These samples were prepared by weighing the samples 10 mg each and transfer in 10 ml volumetric flask. Then, addition of 2 ml HPLC grade methanol in it and dissolve completely. After complete dissolution add sodium hydroxide (0.1 N) and adjust volume 10 ml with it. After complete preparation of the solution, store it at 80°C about 3 h in a water bath (Figure 5).

Oxidation

These samples were prepared by weighing the samples 10 mg each and transfer in 10 ml volumetric flask. Then, addition of 2 ml HPLC grade methanol in it and dissolve completely. After complete dissolution add hydrogen peroxide (3%) and adjust volume 10 ml with it. After complete preparation of the solution, store it at 80°C about 3 h in a water bath (Figure 6).

Thermal degradation

These samples were prepared by weighing the samples 10 mg each and transfer in 10 ml volumetric flask. Then, addition of 2 ml HPLC grade methanol in it and dissolve completely. After complete dissolution add HPLC grade water and adjust volume 10 ml with it. After complete preparation of the solution, store it at 80°C about 3 h in a water bath (Figure 7).
After degrading the samples, allow it to cool. Take 0.5 ml of solution from it and transfer to 10 ml volumetric flask. Adjust volume up to 10 ml with mobile phase. Each sample filtered through 0.22 μ Millipore filter paper for injecting into HPLC.

Method validation

Linearity

Linearity was performed by preparing stock (1000 μg/ml) first then prepare working standard (100 μg/ml) from the stock solution. The linear concentration was prepared and injected in HPLC and response was measured at 274 nm (Figures 8 and 9). The samples injected in triplicates and measure the average peak area to determine standard deviation and relative standard deviation. Furthermore, regression equation was calculated with the help of excel graph by plotting concentration versus peak area (Tables 2 and 3).

Range

The range can be determined as the upper limit of concentration and lower limit of concentration. The linearity range occurs in between 10 and 50 μg/ml. Therefore, the upper concentration limit is 50 μg/ml, and lower concentration limit is 10 μg/ml.


f48f950a-a204-4bf8-9a74-d424ba412137.png

Accuracy

Accuracy was performed with the help of recovery method by standard addition of standard solution in pre-analyzed tablet solution in different levels (80%, 100%, and 120%). Percent recovery was calculated by comparing the area before and after the addition of the standard drug. The solutions were analyzed in triplicate at each level as per the proposed method. The percent recovery and % RSD at each level was calculated and presented in following Table 4. The result indicates that the proposed method was accurate (Tables 4 and 5).

Precision

The precision was performed by intraday (repeatability) and interday (reproducibility) study. The detailed observation is shown in Tables 6-9.

Limit of detection (LOD) and limit of quantitation(LOQ)

LOD is the lowest amount of analyte in a sample that can be detected but not necessarily quantify under the stated experimental conditions. LOQ is the lowest concentration of an analyte in a sample that can be determined with the acceptable precision and accuracy under stated experimental conditions (Table 10).

Robustness

To determine robustness of the method, experimental conditions were purposely altered and chromatographic resolution between atenolol and losartan K were evaluated. The flow rate of the mobile phase was 0.9 ml/min. To study the effect of flow rate on resolution of atenolol and losartan K, it was changed to 0.1 units from 0.8 ml/min to 1 ml/min. While other mobile phase components were held constant. The effect of mobile phase composition on resolution of atenolol and losartan K was studied with methanol: 0.1% O-phosphoric acid at 64:36 (v/v) and 66:34 (v/v). The effect of wavelength on resolution of atenolol and losartan K was studied with applying wavelengths 273 nm and 275 nm. Detailed information about the robustness incorporated inside (Tables 11 and 12).

System suitability

The system suitability parameter such as capacity factor, asymmetry factor, tailing factor, HETP, and number of theoretical plates was also calculated. It was observed that all the values are within the limits and the results are shown in Table 12. The statistical evaluation of the proposed method was revealed its good linearity, reproducibility and its validation for different parameters and let us to the conclusion that it could be used for the rapid and reliable determination of losartan potassium and atenolol in tablet formulation. The results are furnished in Table 13.

Specificity

The specificity of method was performed by comparing the chromatograms of blank, standard and sample. It was found that there is no interference due to excipients in the tablet formulation and also found a good correlation between the retention time of standard and sample of losartan potassium and atenolol. The detail observation is shown in Table 14.


9e80ee4d-ddcc-43d7-85f3-5824291337c0.png


5536594d-a6cf-4176-a076-4e0bf1987fc7.png


9f00d1e5-54d0-4c7f-87c2-8a6615302700.png


2da64ac3-37f5-4694-8218-e850a40847aa.png


a4f495c9-cf88-4d58-be34-b21a279d79a3.png


1353f57f-5ecf-48a4-930b-ca2a9c2d9446.png

Results and Discussion

This study was aimed at stability indicating RP-HPLC method development and validation for simultaneous estimation of losartan potassium, atenolol and their degradation products. A nonpolar C-18 analytical chromatographic column was chosen as the stationary phase for the separation and simultaneous determination of losartan potassium, atenolol and their degradation products. Mixtures of commonly used solvents such as water and methanol in different combinations were tested as mobile phases. The choice of the optimum composition is based on the chromatographic response factor, a good peak shape with minimum tailing. A mixture of methanol and 0.1% O-phosphoric acid in the ratio of 60:35 v/v was proved to be the most suitable of all the combinations since the chromatographic peak obtained was well defined, better resolved and almost free from tailing. The retention times of the losartan potassium and atenolol were found to be 4.05 and 7.76 min, respectively. The forced degradation study was conducted for determining the stability indicating power of an analytical procedure. The result of the stress studies shown in Table 1.
The linearity was found satisfactory for losartan potassium and atenolol in the range 10-50 μg/ml for both Tables 2 and 3. The regression equation of the linearity curve between concentrations of losartan potassium and atenolol over its peak areas were found to be y = 9.152×-8.289 (where “Y” is the peak area and X is the concentration of losartan potassium in μg/ml) and y = 3.816×+1.82 (where Y is the peak area and X is the concentration of atenolol in μg/ml), respectively. The percent recoveries of the drug solutions were studied at three different concentration levels. The percent individual recovery and the %RSD at each level were within the acceptable limits (Tables 4 and 5). This indicates that the method is accurate. The absence of additional peaks in the chromatogram indicates non-interference of the commonly used excipients in the tablets, and hence the method is specific. Precision of the method was studied by repeated and shown (Tables 6-9). This reveals that the method is quite precise. The lowest values of LOD and LOQ as obtained by the proposed method indicate that the method is sensitive (Table 10). The deliberate changes in the method have not much affected the peak tailing, theoretical plates and the percent assay. This indicates that this method is robust (Tables 11 and 12). The system suitability studies were conducted to check various parameters such as theoretical plates and tailing factor (Table 13). The specificity study is shown in Table 14.
Therefore, the proposed method was simple, specific and sensitive and can be used for simultaneous analysis of losartan potassium, atenolol and their degradation products in bulk samples and its tablet dosage forms.

Conclusion

A new analytical method was developed to be routinely applied to simultaneous determination of atenolol and losartan K in pharmaceutical dosage form. In this study, the stability of atenolol and losartan K in present dosage forms was established through employment of ICH recommended stress conditions. The developed procedure was evaluated for specificity, linearity, accuracy, precision, and robustness to ascertain the stability of the analytical method. The method was proved to be specific, linear, precise, accurate, robust, and stability indicating. Hence, the method is recommended for routine quality control analysis and stability sample analysis.

References
  1. Carstensen JT, Rhodes CT. Drug Stability, Principles and Practices. 3rd ed., Vol. 107. 1: Informa Healthcare; 2000. p. 329.
  2. Baertschi SW, Reynolds DW. Pharmaceutical Stress Testing. U.S.A: Taylor & Francis Group; 2005.
  3. Prabhakar AH, Giridhar R. A rapid colorimetric method for the determination of Losartan potassium in bulk and in synthetic mixture for solid dosage form. J Pharm Biomed Anal 2002;27:861-6.
  4. Goodman LS, Gilman A. Diuretics. In: Goodman LS, Hardman JG, Limbird LE, editors. The Pharmacological Basis of Therapeutics. Vol. 10. New York: McGraw Hill; 1986. p. 773-84.
  5. Indian Pharmacopoeia Commission. Indian Pharmacopoeia. Gaziabad: Indian Pharmacopoeia Commission; 2007. p. 701-3.
  6. Abdussalee MK, Boopathy D, Perumal P. Analytical method development and validation of losartan potassium and atenolol in combined dosage form by RP-HPLC. Int J PharmTech Res 2010;1:471-4.
  7. Belal F, Sharaf EM, Aly F, Hefnawy M. Stability-indicating HPLC method for the determination of atenolol in pharmaceutical preparations. J Chromatogr Sep Tech 2013;4:1,1-7.
  8. Seburg RA, Ballard JM, Hwang TL, Sullivan CM. Photosensitized degradation of losartan potassium in an extemporaneous suspension formulation. J Pharm Biomed Anal 2006;42:411-22.
  9. Hertzog DL, McCafferty JF, Fang X, Tyrrell RJ, Reed RA. Development and validation of a stability-indicating HPLC method for the simultaneous determination of Losartan potassium, hydrochlorot hiazide, and their degradation products. J Pharm Biomed Anal 2002;30:747-60.
  10. Patel ND, Captain AD. Extractive spectrophotometric method for simultaneous determination of losartan potassium and atenolol in bulk and in pharmaceutical dosage form. Int J PharmTech Res 2013;5:629-40.
  11. ICH. Harmonized Tripartite Guidelines. Validation of Analytical Procedures: Methodology (Q2B). 1995: ICH; 1996.
  12. ICH. Harmonized Tripartite Guidelines. Text on validation of analytical procedures. Q2A. 1994: ICH; 1994.
  13. ICH. Harmonized Tripartite Guideline. Stability Testing of New Drug Substances and Products Q1A (R2). 2005: ICH; 2003.
  14. ICH - Guidelines Q2A, Validation of Analytical Procedures: Definition and Terminology (CPMP III/5626/94) March (1995). Geneva, Switzerland: ICH. ICH - Guidelines Q2B, Validation of Analytical Procedures: Methodology, (CPMP/ICH/281/95) November (1996). Geneva, Switzerland: ICH.
  15. Carstensen JT, Rhodes CT. Drug Stability Principles and Practices. 3rd ed. New York: Dekkar Series; 2000.